Title:
Enhanced electrochemical biosensing efficiency of silica particles supported on partially reduced graphene oxide for sensitive detection of cholesterol

dc.contributor.authorShiju Abraham
dc.contributor.authorSaurabh Srivastava
dc.contributor.authorVinod Kumar
dc.contributor.authorShobhit Pandey
dc.contributor.authorPankaj Kumar Rastogi
dc.contributor.authorNarsingh R. Nirala
dc.contributor.authorSunayana Kashyap
dc.contributor.authorSunil K. Srivastava
dc.contributor.authorVidya Nand Singh
dc.contributor.authorVellaichamy Ganesan
dc.contributor.authorPreeti S. Saxena
dc.contributor.authorAnchal Srivastava
dc.date.accessioned2026-02-07T06:09:55Z
dc.date.issued2015
dc.description.abstractThe present work introduces partially reduced graphene oxide (pRGO)-silica (SiO2) particles hybrid system (pRGOSHs) for sensitive and cost effective free cholesterol detection. Fabricated out of thin layers of pRGOSHs, these proposed ChOx/pRGOSHs/ITO based biosensors have a detection range of 2.6-15.5 mM with an appreciable detection limit of 1.3 mM and sensitivity of 11.1 μA/mM/cm2. Low Michaelis-Menten constant (Km) (4.9 × 10- 4 mM) and high diffusivity constant (D) (3.2 × 10- 10 cm2/s) values clearly indicate enhanced immobilization of enzyme over the substrate. Additionally, electrochemical impedance studies indicate that the synergistic combination of SiO2 and pRGO also results in much lower impedance values (40% and 18% decrease in comparison to SiO2 and pRGO respectively) for an overall enhanced sensing performance. These results are further corroborated by the density functional theory based theoretical simulations indicating enhanced electron density (theoretically) in case of the proposed hybrid system. Finally, the present work also highlights the importance of Si-OH bonds formation in the proposed pRGOSHs composite system for attaining such enhanced biosensing ability. © 2015 Elsevier B.V.
dc.identifier.doi10.1016/j.jelechem.2015.09.016
dc.identifier.issn15726657
dc.identifier.urihttps://doi.org/10.1016/j.jelechem.2015.09.016
dc.identifier.urihttps://dl.bhu.ac.in/bhuir/handle/123456789/27344
dc.publisherElsevier B.V.
dc.subjectCyclic voltammetry
dc.subjectK<sub>m</sub> value
dc.subjectPartially reduced graphene oxide, cholesterol sensing
dc.subjectSilica particles
dc.titleEnhanced electrochemical biosensing efficiency of silica particles supported on partially reduced graphene oxide for sensitive detection of cholesterol